Repository logo
 

The development and optimisation of 3D black-blood R2 * mapping of the carotid artery wall

Accepted version
Peer-reviewed

No Thumbnail Available

Type

Article

Change log

Authors

Yuan, J 
Patterson, A 
Priest, A 

Abstract

Purpose To develop and optimise a 3D black-blood R2* mapping sequence for imaging the carotid artery wall, using optimal blood suppression and k-space view ordering.

Methods Two different blood suppression preparation methods were used; Delay Alternating with Nutation for Tailored Excitation (DANTE) and improved Motion Sensitive Driven Equilibrium (iMSDE) were each combined with a three-dimensional (3D) multi-echo Fast Spoiled GRadient echo (ME-FSPGR) readout. Three different k-space view-order designs: Radial Fan-beam Encoding Ordering (RFEO), Distance-Determined Encoding Ordering (DDEO) and Centric Phase Encoding Order (CPEO) were investigated. The sequences were evaluated through Bloch simulation and in a cohort of twenty volunteers. The vessel wall Signal-to-Noise Ratio (SNR), Contrast-to-Noise Ratio (CNR) and R2*, and the sternocleidomastoid muscle R2* were measured and compared. Different numbers of acquisitions-per-shot (APS) were evaluated to further optimise the effectiveness of blood suppression.

Results All sequences resulted in comparable R2* measurements to a conventional, i.e. non-blood suppressed sequence in the sternocleidomastoid muscle of the volunteers. Both Bloch simulations and volunteer data showed that DANTE has a higher signal intensity and results in a higher image SNR than iMSDE. Blood suppression efficiency was not significantly different when using different k-space view orders. Smaller APS achieved better blood suppression.

Conclusion The use of blood-suppression preparation methods does not affect the measurement of R2*. DANTE prepared ME-FSPGR sequence with a small number of acquisitions-per-shot can provide high quality black-blood R2* measurements of the carotid vessel wall.

Description

Keywords

blood suppression, quantitative R2* mapping, carotid imaging

Journal Title

Magnetic Resonance Imaging

Conference Name

Journal ISSN

0730-725X
1873-5894

Volume Title

44

Publisher

Elsevier
Sponsorship
Addenbrooke's Charitable Trust (ACT) (CT/9259)
MRC (1621090)
The project was supported by the Addenbrooke’s Charitable Trust and the National Institute of Health Research Comprehensive Biomedical Research Centre award to Cambridge University Health Partners