Accuracy of four lateral flow immunoassays for anti SARS-CoV-2 antibodies: a head-to-head comparative study

Published version
Peer-reviewed

Type

Article

Change log

Authors

Jones, Hayley 
Mulchandani, Ranya 
Taylor-Phillips, Sian 
Ades, AE 
Shute, Justin 

Abstract

Background

SARS-CoV-2 antibody tests are used for population surveillance and might have a future role in individual risk assessment. Lateral flow immunoassays (LFIAs) can deliver results rapidly and at scale, but have widely varying accuracy.

Methods

In a laboratory setting, we performed head-to-head comparisons of four LFIAs: the Rapid Test Consortium’s AbC-19 ™ Rapid Test, OrientGene COVID IgG/IgM Rapid Test Cassette, SureScreen COVID-19 Rapid Test Cassette, and Biomerica COVID-19 IgG/IgM Rapid Test. We analysed blood samples from 2,847 key workers and 1,995 pre-pandemic blood donors with all four devices.

Findings

We observed a clear trade-off between sensitivity and specificity: the IgG band of the SureScreen device and the AbC-19 ™ device had higher specificities but OrientGene and Biomerica higher sensitivities. Based on analysis of pre-pandemic samples, SureScreen IgG band had the highest specificity (98.9%, 95% confidence interval 98.3 to 99.3%), which translated to the highest positive predictive value across any pre-test probability: for example, 95.1% (95%CI 92.6, 96.8%) at 20% pre-test probability. All four devices showed higher sensitivity at higher antibody concentrations (“spectrum effects”), but the extent of this varied by device.

Interpretation

The estimates of sensitivity and specificity can be used to adjust for test error rates when using these devices to estimate the prevalence of antibody. If tests were used to determine whether an individual has SARS-CoV-2 antibodies, in an example scenario in which 20% of individuals have antibodies we estimate around 5% of positive results on the most specific device would be false positives.

Funding

Public Health England.

Research in context

Evidence before this study

We searched for evidence on the accuracy of the four devices compared in this study: OrientGene COVID IgG/IgM Rapid Test Cassette, SureScreen COVID-19™ Rapid Test Cassette, Biomerica COVID-19 IgG/IgM Rapid Test and the UK Rapid Test Consortium’s AbC-19™ Rapid Test. We searched Ovid MEDLINE (In-Process & Other Non-Indexed Citations and Daily), PubMed, MedRxiv/BioRxiv and Google Scholar from January 2020 to 16 th January 2021. Search terms included device names AND ((SARS-CoV-2) OR (covid)). Of 303 records assessed, data were extracted from 24 studies: 18 reporting on the accuracy of the OrientGene device, 7 SureScreen, 2 AbC-19™ and 1 Biomerica. Only three studies compared the accuracy of two or more of the four devices. With the exception of our previous report on the accuracy of the AbC-19™ device, which the current manuscript builds upon, sample size ranged from 7 to 684. For details, see Supplementary Materials. The largest study compared OrientGene, SureScreen and Biomerica. SureScreen was estimated to have the highest specificity (99.8%, 95% CI 98.9 to 100%) and OrientGene the highest sensitivity (92.6%), but with uncertainty about the latter result due to small sample sizes. The other two comparative studies were small (n = 65, n = 67) and therefore provide very uncertain results. We previously observed spectrum effects for the AbC-19™ device, such that sensitivity is upwardly biased if estimated only from PCR-confirmed cases. The vast majority of previous studies estimated sensitivity in this way.

Added value of this study

We performed a large scale (n = 4,842), head-to-head laboratory-based evaluation and comparison of four lateral flow devices, which were selected for evaluation by the UK Department of Health and Social Care’s New Tests Advisory Group, on the basis of a survey of test and performance data available. We evaluated the performance of diagnosis based on both IgG and IgM bands, and the IgG band alone. We found a clear trade-off between sensitivity and specificity across devices, with the SureScreen and AbC-19™ devices being more specific and OrientGene and Biomerica more sensitive. Based on analysis of 1,995 pre-pandemic blood samples, we are 99% confident that SureScreen (IgG band reading) has the highest specificity of the four devices (98.9%, 95% CI 98.3, 99.3%). We found evidence that all four devices have reduced sensitivity at lower antibody indices, i.e. spectrum effects. However, the extent of this varies by device and appears to be less for other devices than for AbC-19. Our estimates of sensitivity and specificity are likely to be higher than would be observed in real use of these devices, as they were based on majority readings of three trained laboratory personnel.

Implications of all the available evidence

When used in epidemiological studies of antibody prevalence, the estimates of sensitivity and specificity provided in this study can be used to adjust for test errors. Increased precision in error rates will translate to increased precision in seroprevalence estimates. If lateral flow devices were used for individual risk assessment, devices with maximum specificity would be preferable. However, if, for an example, 20% of the tested population had antibodies, we estimate that around 1 in 20 positive results on the most specific device would be incorrect.

Description

Keywords

COMPARE study investigators, EDSAB-HOME investigators

Journal Title

EBioMedicine

Conference Name

Journal ISSN

2352-3964
2352-3964

Volume Title

68

Publisher

Elsevier
Sponsorship
Cambridge University Hospitals NHS Foundation Trust (CUH) (unknown)
British Heart Foundation (CH/12/2/29428)
British Heart Foundation (None)
British Heart Foundation (RG/18/13/33946)
Medical Research Council (MR/L003120/1)
Relationships
Is derived from: