Simulation method for investigating the use of transition-edge sensors as spectroscopic electron detectors
Publication Date
2021Journal Title
Superconductor Science and Technology
ISSN
0953-2048
Publisher
IOP Publishing
Volume
34
Issue
12
Language
en
Type
Article
This Version
VoR
Metadata
Show full item recordCitation
Patel, K., Withington, S., Thomas, C., Goldie, D., & Shard, A. (2021). Simulation method for investigating the use of transition-edge sensors as spectroscopic electron detectors. Superconductor Science and Technology, 34 (12) https://doi.org/10.1088/1361-6668/ac30d0
Abstract
Transition-edge sensors (TESs) are capable of highly accurate single particle
energy measurement. TESs have been used for a wide range of photon detection
applications, particularly in astronomy, but very little consideration has been
given to their capabilities as electron calorimeters. Existing electron
spectrometers require electron filtering optics to achieve energy
discrimination, but this step discards the vast majority of electrons entering
the instrument. TESs require no such energy filtering, meaning they could
provide orders of magnitude improvement in measurement rate. To investigate the
capabilities of TESs in electron spectroscopy, a simulation pipeline has been
devised. The pipeline allows the results of a simulated experiment to be
compared with the actual spectrum of the incident beam, thereby allowing
measurement accuracy and efficiency to be studied. Using Fisher information,
the energy resolution of the simulated detectors was also calculated, allowing
the intrinsic limitations of the detector to be separated from the specific
data analysis method used. The simulation platform has been used to compare the
performance of TESs with existing X-ray photoelectron spectroscopy (XPS)
analysers. TESs cannot match the energy resolution of XPS analysers for
high-precision measurements but have comparable or better resolutions for high
count rate applications. The measurement rate of a typical XPS analyser can be
matched by an array of 10 TESs with 120 microsecond response times and there is
significant scope for improvement, without compromising energy resolution, by
increasing array size.
Keywords
transition-edge sensors, electron spectroscopy, x-ray photoelectron spectroscopy, superconducting detectors, calorimeter, simulation
Sponsorship
Engineering and Physical Sciences Research Council (EP/L015978/1)
Identifiers
sustac30d0, ac30d0, sust-104598.r1
External DOI: https://doi.org/10.1088/1361-6668/ac30d0
This record's URL: https://www.repository.cam.ac.uk/handle/1810/330138
Rights
Licence:
http://creativecommons.org/licenses/by/4.0
Statistics
Total file downloads (since January 2020). For more information on metrics see the
IRUS guide.
Recommended or similar items
The current recommendation prototype on the Apollo Repository will be turned off on 03 February 2023. Although the pilot has been fruitful for both parties, the service provider IKVA is focusing on horizon scanning products and so the recommender service can no longer be supported. We recognise the importance of recommender services in supporting research discovery and are evaluating offerings from other service providers. If you would like to offer feedback on this decision please contact us on: support@repository.cam.ac.uk