Repository logo
 

Stability of Thin Film Neuromodulation Electrodes under Accelerated Aging Conditions

Accepted version
Peer-reviewed

Loading...
Thumbnail Image

Change log

Abstract

Thin film electrodes are becoming increasingly common for interfacing with tissue. However, their long-term stability has yet to be proven in neuromodulation applications where electrical stimulation over months to years is desired. Here, we examine the stability of pristine and PEDOT:PSS-coated Au, as well as pristine PEDOT:PSS microelectrodes over a period of three months in an accelerated aging set up where they are exposed to current stimulation, hydrogen peroxide, mechanical agitation and high temperature. Pristine PEDOT:PSS electrodes show the highest stability, while pristine Au electrodes show the lowest stability. Failure mode analysis reveals that delamination and Au corrosion are the key drivers of electrode degradation. The PEDOT:PSS coating slows down Au corrosion to a degree that depends on the overlap between the two films. The results demonstrate that pristine PEDOT:PSS electrodes represent a promising way forward towards thin film devices for long-term in vivo neuromodulation applications.

Description

Journal Title

Advanced Functional Materials

Conference Name

Journal ISSN

1616-301X
1616-3028

Volume Title

Publisher

Wiley

Rights and licensing

Except where otherwised noted, this item's license is described as All Rights Reserved