Repository logo
 

White light polarization sensitive optical coherence tomography for sub-micron axial resolution and spectroscopic contrast in the murine retina.

Published version
Peer-reviewed

Repository DOI


Loading...
Thumbnail Image

Change log

Abstract

A white light polarization sensitive optical coherence tomography system has been developed, using a supercontinuum laser as the light source. By detecting backscattered light from 400 - 700 nm, an axial resolution of 1.0 µm in air was achieved. The system consists of a free-space interferometer and two homemade spectrometers that detect orthogonal polarization states. Following system specifications, images of a healthy murine retina as acquired by this non-contact system are presented, showing high resolution reflectivity images as well as spectroscopic and polarization sensitive contrast. Additional images of the very-low-density-lipoprotein-receptor (VLDLR) knockout mouse model were acquired. The high resolution allows the detection of small lesions in the retina.

Description

Journal Title

Biomed Opt Express

Conference Name

Journal ISSN

2156-7085
2156-7085

Volume Title

9

Publisher

Optica Publishing Group

Rights and licensing

Except where otherwised noted, this item's license is described as Attribution 4.0 International (CC BY 4.0)
Sponsorship
European Research Council
FWF Austrian Science Fund